User interface | LCD monochrome dot matrix, 240 x 128, white LED backlight; 5 button keypad |
Display capability | Up to eight simultaneous channels of pressure measurement, up to a 5 character user label for each; up to two channels in a large font; readable from 15 feet; display updates every 200 msec |
Gauge capability | Up to 4 Hot Filament BA gauges including metal, glass, and nude UHV types using the HFIG** cards; up to 5 IMGs* using the IMG* cards; up to 12 rough gauges using the Dual Convection Gauge Cards; up to 12 Active Gauges using Dual Analog Input Cards |
Measurement range | 1 x 10-11 Torr to Atm dependent upon card and gauge type selected |
Setpoints | Eight normally open setpoints (open collector), with independently programmable ON and OFF delays,manual override (see detail specs for ratings) |
Fast setpoint response | 20 msec max (with delay set to 0.0 secs) |
Serial communications | RS232 and RS485 (fully programmable), Profibus DP (optional) |
Analog outputs | 1v/dec LOG, 0 – 10 V FS, on each card with 200 msec update rate |
I/O control | HFIG** card: Remote gauge ON/OFF, Fil select, DeGAS ON/OFF Inputs, Status Output |
| IMG* card: Remote gauge ON/OFF Input, Status Output |
| Dual Convection Card: Remote CAL inputs, Status Outputs |
| Dual Analog Input Card: Remote CAL Inputs, Status Output |
Other | Auto-ON capability for installed ion gauges when at least one convection card is installed |
| Sensitivity, gas correction, emission current (BA gauges only) inputs for ion gauges |
| E-BEAM degas, with auto turn off for BA gauges |
| Selectable air or argon calibration for rough gauges |
| Adjustable atmosphere value for rough gauges |
| Text error messages |
| All data stored in FLASH non-volatile memory requiring no batteries |
Power input/Temperature | Worldwide universal power input, 100-240 VAC, 2 A, 50/60 Hz, IEC inlet; 5-40 °C ambient |
Regulatory approvals | CE certified and RoHS compliant, BV certified for USA and Canada |